The ER-1250S/SR-2790C/SR-2108A infrared blackbody cavity produced by AIKOM company, its are designed and manufactured according to the technical standards of "ASTM E1965-98 (2009)" and national calibration specifications of "JJF-1107-2003" and "JJF1577-2016". These three infrared blackbody cavity can be used in conjunction with our company's constant temperature water baths for standard transmission of infrared ear/forehead thermometers and calibration of industrial radiation thermometers.

All Video Query


  • Product Description
  • Application areas
  • Product characteristics
  • Specificatiom
  • Product Warranty
  • After-sale service
  • Environmental impact
  • optional accessories
  • Product Description
    • Commodity name: Infrared ear thermometer blackbody cavity
    • Commodity ID: ER-1250S

    The ER-1250S/SR-2790C/SR-2108A infrared blackbody cavity produced by AIKOM company, its are designed and manufactured according to the technical standards of "ASTM E1965-98 (2009)" and national calibration specifications of "JJF-1107-2003" and "JJF1577-2016". These three infrared blackbody cavity can be used in conjunction with our company's constant temperature water baths for standard transmission of infrared ear/forehead thermometers and calibration of industrial radiation thermometers.

    The ER-1250S/SR-2790C/SR-2108A infrared blackbody cavity produced by AIKOM company, its are designed and manufactured according to the technical standards of "ASTM E1965-98 (2009)" and national calibration specifications of "JJF-1107-2003" and "JJF1577-2016". These three infrared blackbody cavity can be used in conjunction with our company's constant temperature water baths for standard transmission of infrared ear/forehead thermometers and calibration of industrial radiation thermometers.  

    Key words:
    • Infrared blackbody
  • Application areas

    Used for traceability of human ear therm-ometer measurement and calibration, as well as research and development, calibration, testing and calibration of infrared ear thermometer manufacturers.

  • Product characteristics

    ●Processed from the entire copper material, without any external coating, with good conductivity to ensure the emissivity of the blackbody cavity.
    ●Completely submerged, equipped with anti floating swing card slots (matched with our company's constant temperature bath series products).
    ●Equipped with a standard thermometer mounting bracket. Convenient for users to use a constant temperature bath as a temperature source for the blackbody cavity.
    ●Equipped with dust covers and instrument boxes.

  • Specificatiom

    Temperature range:  20~50℃
    Recurrent:  8-14UMbrightness temp.
    Effective emissivity:  ≥0.9995
    Diameter of cavity opening:  Φ8-10mm (non concentric)
    Inner diameter of the cavity:  About 60mm
    Cavity depth:  About 118mm
    Excessive disc diameter:  About 126mm
    Standard thermometer socket diameter:  About 7.8mm (6.5mm)